書誌事項
- タイトル別名
-
- Inter-laboratory Comparison of the Co-C Eutectic Point with Small-type Cell for Thermocouples
- ネツデンツイヨウ コガタ キョウショウテン セル ノ モチマワリ シケン
この論文をさがす
抄録
Using a small-type crucible, the inter-laboratory comparison of the realization of the Co-C eutectic point was executed in the industrial field in Japan, with the aim to verify the possibility of thermocouple calibration at the eutectic point with an existing furnace of not so good temperature distribution: the palladium-point wire-bridge method furnace for thermocouple calibration. A small-type crucible for thermocouples was made by the National Metrology Institute of Japan, National Institute of Advanced Industrial Science and Technology (NMIJ/AIST). The inter-laboratory comparison was piloted by the Tokyo Metropolitan Industrial Technology Research Institute (TIRI), a public test and research organization in Tokyo, and was conducted among five calibration laboratories including the pilot. One of the five laboratories is a designated calibration laboratory in Japan, and the three remainders are private laboratories. Four laboratories including TIRI used existing palladium-point furnace, and one laboratory made a new furnace for this inter-laboratory comparison.
収録刊行物
-
- 電気学会論文誌. A
-
電気学会論文誌. A 132 (12), 1126-1131, 2012
一般社団法人 電気学会
- Tweet
詳細情報 詳細情報について
-
- CRID
- 1390282679577108096
-
- NII論文ID
- 10031129647
-
- NII書誌ID
- AN10136312
-
- ISSN
- 13475533
- 03854205
-
- NDL書誌ID
- 024252665
-
- 本文言語コード
- ja
-
- データソース種別
-
- JaLC
- NDL
- Crossref
- CiNii Articles
-
- 抄録ライセンスフラグ
- 使用不可