植込み型除細動器への実装を考慮した致死性不整脈検出アルゴリズムの改良

書誌事項

タイトル別名
  • Improving Detection Algorithm of Life-threatening Arrhythmias for Implementation of Implantable Cardioverter-Defibrillators
  • ウエコミ ガタジョサイドウキ エ ノ ジッソウ オ コウリョ シタ チシセイ フセイミャク ケンシュツ アルゴリズム ノ カイリョウ

この論文をさがす

抄録

The implantable cardioverter-defibrillator (ICD) is an effective therapeutic device for rescuing patients with cardiac diseases from death caused by life-threatening arrhythmias. The authors previously proposed a detection algorithm of life-threatening arrhythmias with a multiple regression model. To enhance the classification accuracy, in the present study, we have introduced an autoregressive filter and a multiple detection process into the previous detection algorithm. The experimental results showed that the proposed method could attain a high accuracy such that all ventricular fibrillation rhythms could be exactly detected. In addition, detection errors of sinus rhythms or supraventricular tachyarrhythmias provoking the ICD malfunction were reduced.