書誌事項
- タイトル別名
-
- Tissue Extraction of Ultrasonic Image by Using Two-Dimensional AR Model
- 2ジゲンバ ジコ カイキ モデル オ モチイタ チョウオンパ ガゾウ ニ オケ
この論文をさがす
説明
An image obtained by ultrasonic medical equipment is poor in quality because of speckle noise, that is caused by the quality of ultrasonic beam and so on. Thus, it is very difficult to detect internal organs or the diseased tissues from a medical ultrasonic image by the processing, which is used only gray-scale of the image. To analyze the ultrasonic image, it is necessary to use not only gray-scale but also appropriate statistical character. In this paper, we suggest a new method to extract regions of internal organs from an ultrasonic image by the discrimination function. The discrimination function is based on gray-scale and statistical characters of the image. This function is determined by using parameters of the multi-dimensional autoregressive model.
収録刊行物
-
- 電気学会論文誌C(電子・情報・システム部門誌)
-
電気学会論文誌C(電子・情報・システム部門誌) 115 (12), 1444-1451, 1995
一般社団法人 電気学会
- Tweet
キーワード
詳細情報 詳細情報について
-
- CRID
- 1390282679585698176
-
- NII論文ID
- 130006844531
- 10004156721
- 10001708851
-
- NII書誌ID
- AN10065950
-
- ISSN
- 13488155
- 03854221
-
- NDL書誌ID
- 3634697
-
- データソース種別
-
- JaLC
- NDLサーチ
- Crossref
- CiNii Articles
- OpenAIRE
-
- 抄録ライセンスフラグ
- 使用不可