M系列符号化積層探触子によるパルス圧縮超音波撮像に関する研究

書誌事項

タイトル別名
  • Study on Pulse Compression Ultrasonic Imaging by a Stacked M-sequence Encoding Probe
  • M ケイレツ フゴウカ セキソウタンショクコ ニ ヨル パルス アッシュク チョウオンパ サツゾウ ニ カンスル ケンキュウ

この論文をさがす

抄録

<p>If an ultrasonic imaging inside an object is possible without scanning by using an ultrasonic probe having a small aperture, the ultrasonic imaging system becomes a simpler and easier system. We have developed a stacked encoding ultrasonic transducer which can transceive an M-sequence signal only by a pulse drive in order to realize an ultrasonic measurement with a high resolution and high sensitivity. When an ultrasonic probe with such transducer is used, a pulse compression measurement is enabled only with a general ultrasonic flaw detector and a PC without using special devices. In this study, we focused on the characteristic that the received waveform by this probe changes with transceiver directions of a wave. A stacked M-sequence encoding probe of 7 bits length was manufactured and the one-dimensional point spread function in transceiver angle was obtained. As a result, it was confirmed that this probe has not only the pulse compression characteristic but the angle selecting characteristic. Finally, by using the proposed probe, the cross-sectional imaging for point objects without scanning was demonstrated.</p>

収録刊行物

参考文献 (2)*注記

もっと見る

詳細情報 詳細情報について

問題の指摘

ページトップへ