書誌事項
- タイトル別名
-
- Reduction of Environmental Magnetic Field Noise for Biomagnetic Measurements
- セイタイ ジキ ケイソク ノ タメ ノ テイジバ クウカン ノ セイセイ
この論文をさがす
抄録
In order to measure biomagnetic fields such as magnetoencephalogram (MEG), magnetically shielded room (MSR) with multi layers of μ-metal is usually used to reduce environmental magnetic field noise. However, such heavy-shielded room is not suitable for magnetocardiogram (MCG) measurement system from the requirements of low cost, lightweight, and small space.<br> We have studied an environmental noise reduction scheme for MCG application. Our noise reduction system consists of a prefabricated MSR of single μ-metal layer, a noise detection coil, and compensation coils. The following results were obtained for the suppression of environmental noise: 30 dB at 0.5-10 Hz and 30-50 dB at 10-100 Hz.<br> In this paper, the compensation set-ups are described and experimental results are presented and discussed.
収録刊行物
-
- 電気学会論文誌C(電子・情報・システム部門誌)
-
電気学会論文誌C(電子・情報・システム部門誌) 121 (11), 1704-1710, 2001
一般社団法人 電気学会
- Tweet
詳細情報 詳細情報について
-
- CRID
- 1390282679587187712
-
- NII論文ID
- 130006845499
- 10007233819
-
- NII書誌ID
- AN10065950
-
- ISSN
- 13488155
- 03854221
-
- NDL書誌ID
- 5962835
-
- データソース種別
-
- JaLC
- NDL
- Crossref
- CiNii Articles
-
- 抄録ライセンスフラグ
- 使用不可