- 榊原 潤
- 筑波大学構造工学系
書誌事項
- タイトル別名
-
- Measurment of whole field temperature using two-color LIF
- 2ショク LIF ニヨル 3ジゲン オンド ケイソクホウ ノ カイハツ
この論文をさがす
抄録
A technique to measure the instantaneous three-dimensional temperature distribution of water was established using two-color laser induced fluorescence (LIF). Ratio of fluorescence intensity of two fluorescence dye was calibrated against the temperature to eliminate the effect of the fluctuation of exciting light intensity. Stable thermally stratified layer was measured by this system to evaluate the total accuracy of the measurement system. A random error of the measurement was±1.4K with 95% confidence. Measurement of an unsteady penetrative convection showed typical structures e.g. plumes and ridges in temperature iso-surfaces and contours.
収録刊行物
-
- 可視化情報学会誌
-
可視化情報学会誌 17 (Supplement1), 333-336, 1997
社団法人 可視化情報学会
- Tweet
詳細情報 詳細情報について
-
- CRID
- 1390282679597096320
-
- NII論文ID
- 10002674948
- 130003648706
-
- NII書誌ID
- AN10374478
-
- ISSN
- 1884037X
- 09164731
-
- NDL書誌ID
- 4247100
-
- 本文言語コード
- ja
-
- データソース種別
-
- JaLC
- NDL
- Crossref
- CiNii Articles
-
- 抄録ライセンスフラグ
- 使用不可