書誌事項
- タイトル別名
-
- Design Method of Digital Optimal Control Scheme and Multiple Paralleled Bridge Type Current Amplifier for Generating Gradient Magnetic Fields in MRI Systems
- MRI ケイシャ ジバ ハッセイヨウ タヘイレツ ブリッジ デンリュウ ゾウフクキ ト ディジタル サイテキ セイギョ セッケイホウ
この論文をさがす
説明
This paper deals with a digital control scheme of multiple paralleled high frequency switching current amplifier with four-quadrant chopper for generating gradient magnetic fields in MRI (Magnetic Resonance Imaging) systems. In order to track high precise current pattern in Gradient Coils (GC), the proposal current amplifier cancels the switching current ripples in GC with each other and designed optimum switching gate pulse patterns without influences of the large filter current ripple amplitude. The optimal control implementation and the linear control theory in GC current amplifiers have affinity to each other with excellent characteristics. The digital control system can be realized easily through the digital control implementation, DSPs or microprocessors. Multiple-parallel operational microprocessors realize two or higher paralleled GC current pattern tracking amplifier with optimal control design and excellent results are given for improving the image quality of MRI systems.
収録刊行物
-
- 電気学会論文誌D(産業応用部門誌)
-
電気学会論文誌D(産業応用部門誌) 123 (8), 894-902, 2003
一般社団法人 電気学会
- Tweet
詳細情報 詳細情報について
-
- CRID
- 1390282679633495936
-
- NII論文ID
- 10011450437
-
- NII書誌ID
- AN10012320
-