Determination of Copper(II) in Concentrated Alkaline Solutions by Anodic Stripping Voltammetry with a Microporous Membrane Electrode

Bibliographic Information

Other Title
  • 多孔質膜金電極における陽極溶出ボルタンメトリーによる高濃度アルカリ水溶液中の微量銅イオンの定量
  • タコウシツマクキン デンキョク ニ オケル ヨウキョク ヨウシュツボルタンメトリー ニ ヨル コウノウド アルカリ スイヨウエキ チュウ ノ ビリョウ ドウ イオン ノ テイリョウ

Search this article

Abstract

A new anodic stripping voltammetry (ASV) using a microporous membrane electrode has been developed for the on-line monitoring of trace CuII in concentrated alkaline solutions. Microporous membrane electrodes were prepared by plating gold onto membrane filters made of polypropylene (PP). By SEM observations, it was confirmed that the membrane filter was modified by gold fine particles. The microporous membrane electrodes were set in a penetrated-type electrochemical flow-cell, and CuII was deposited while the sample solution passed through the electrode. The concentration of CuII was estimated with the anodic stripping peak currents. The calibration curve showed linearity within a range of 0 – 25 μg L−1 CuII in a 48% KOH solution. The proposed method will become a useful technique for an on-line monitoring system for trace CuII, which has been in urgent demand by the semiconductor manufacturing.

Journal

  • BUNSEKI KAGAKU

    BUNSEKI KAGAKU 62 (8), 707-712, 2013