- DOI Koichiro
- Junior College of Engineering, Shizuoka University
-
- SENNO Toshiyuki
- Material Department of Toyo Kogyo Co., Ltd.
-
- MATSUI Yoshikazu
- Faculty of Engineering, Shizuoka University
Bibliographic Information
- Other Title
-
- 可動管質量流量計
- カドウカン シツリョウ リュウリョウケイ
Search this article
Abstract
The movable tube mass flowmeter here reported consists of two movable tube flowmeters and two constant flow apparatus. This mass flowmeter has such a merit as the mass flow rates can be measured regardless of the viscosity of the fluid to be measured. This merit is very important because a change in density is frequently accompanied with a change in viscosity.<br> The present movable tube mass flowmeter may be classified into two types, namely, with and without a pressure compensator. By the latter, the mass flow rate is calculated from three measured values and by the former, from only one measured value.<br> The experimental results using trial mass flowmeters of both the above mentioned types, agree fairly well with theory.
Journal
-
- Oyo Buturi
-
Oyo Buturi 42 (11), 1086-1090, 1973
The Japan Society of Applied Physics
- Tweet
Details 詳細情報について
-
- CRID
- 1390282679639966464
-
- NII Article ID
- 130003589625
-
- NII Book ID
- AN00026679
-
- ISSN
- 21882290
- 03698009
-
- NDL BIB ID
- 7503533
-
- Data Source
-
- JaLC
- NDL
- CiNii Articles
-
- Abstract License Flag
- Disallowed