A case of high venous pressure caused by buttonhole puncture using a painless needle with side lumens
-
- Hayama Atsuko
- Department of Nursing, Fukuyama City Hospital
-
- Kaihara Keiko
- Department of Nursing, Fukuyama City Hospital
-
- Yamauchi Chiduru
- Department of Nursing, Fukuyama City Hospital
-
- Okamoto Kayo
- Department of Nursing, Fukuyama City Hospital
-
- Takahashi Tomoko
- Department of Nursing, Fukuyama City Hospital
-
- Kusaura Maki
- Department of Nursing, Fukuyama City Hospital
-
- Yamasaki Chizue
- Department of Nursing, Fukuyama City Hospital
-
- Habara Nahomi
- Department of Nursing, Fukuyama City Hospital
-
- Tanaka Shoko
- Department of Medical Engineering, Fukuyama City Hospital
-
- Shimokanekuri Hiroko
- Department of Medical Engineering, Fukuyama City Hospital
-
- Kanemitsu Satomi
- Department of Medical Engineering, Fukuyama City Hospital
-
- Tokura Takehiko
- Department of Internal Medicine, Fukuyama City Hospital
-
- Kurose Yuko
- Department of Internal Medicine, Fukuyama City Hospital
-
- Hirayama Takashi
- Department of Internal Medicine, Fukuyama City Hospital
-
- Sasaki Tamaki
- Division of Nephrology and Hypertension, Department of Internal Medicine, Kawasaki Medical School
Bibliographic Information
- Other Title
-
- 側溝のないペインレスニードル使用によりボタンホール穿刺時の静脈圧上昇を改善しえた1例
- 症例報告 側溝のないペインレスニードル使用によりボタンホール穿刺時の静脈圧上昇を改善しえた1例
- ショウレイ ホウコク ソッコウ ノ ナイ ペインレスニードル シヨウ ニ ヨリ ボタンホール センシジ ノ ジョウミャクアツ ジョウショウ オ カイゼン シエタ 1レイ
Search this article
Description
<p>The buttonhole puncture technique was used in a 70-year-old female dialysis patient to reduce the puncture pain associated with dialysis. However, her venous pressure increased 10 minutes after dialysis was started, and it was speculated that the side lumens of the painless needle used in the procedure had caused the increase in venous pressure. After using the non-side lumens of the painless needle, the patient’s venous pressure fell. This is considered to be an informative case, which will be helpful for facilities that are considering the introduction of the buttonhole puncture technique.</p>
Journal
-
- Nihon Toseki Igakkai Zasshi
-
Nihon Toseki Igakkai Zasshi 51 (4), 289-293, 2018
The Japanese Society for Dialysis Therapy
- Tweet
Details 詳細情報について
-