Development of Velocity Spectrometer for Neutral Atomic Beam

Bibliographic Information

Other Title
  • 中性原子ビーム速度分光装置の開発
  • チュウセイ ゲンシ ビーム ソクド ブンコウ ソウチ ノ カイハツ

Search this article

Abstract

A velocity spectrometer for a neutral atomic beam has been developed to create a mono-velocity neutral atomic beam. In order to disperse a neutral atomic beam depending on its velocity, a magnetic circuit with a large and uniform magnetic gradient has been developed for the first time. The magnetic circuit was made up of quadrupole field configuration of NdFeB permanent magnets. The magnetic gradient in the circuit was measured to be sufficiently large (1T/cm) and uniform. It is possible to use the magnetic circuit as a focusing lens as well as a prism of the spectrometer for a neutral atomic beam. The resonant fluorescence spectra of the Li atomic beam that passed through the magnetic circuit were measured by using a single mode tunable laser. From the peak shift of the fluorescence spectrum, the exerted force on Li atoms by the magnetic gradient in the circuit was 10000 times as large as the earth's gravity. It is expected that a mono-velocity Li atomic beam of 1000m/s with 1% resolution can be created with a compact experimental arrangement using the developed velocity spectrometer.

Journal

References(12)*help

See more

Details 詳細情報について

Report a problem

Back to top