Report of the Committee on Standardization of Laboratory Testing Related to Diabetes Mellitus of the Japan Diabetes Society: Determination of Reference Intervals of Hemoglobin A1C (IFCC) and Glycoalbumin in the Japanese Population
-
- Tominaga Makoto
- Department of Laboratory Medicine, Yamagata University School of Medicine
-
- Makino Hideichi
- Department of Molecular and Genetic Medicine, Ehime University School of Medicine
-
- Yoshino Gen
- Department of Laboratory Medicine, Toho University School of Medicine
-
- Kuwa Katsuhiko
- Institute of Clinical Medicine, University of Tsukuba
-
- Takei Izumi
- Department of Laboratory Medicine, Keio University School of Medicine
-
- Aono Yukuko
- Tokyo Bunka College of Medical Technology
-
- Hoshino Tadao
- Institute of Biopathological Medicine
-
- Umemoto Masao
- HECTEF Standard Reference Center Foundation
-
- Shimatsu Akira
- Clinical Research Institute, National Hospital Organization Kyoto Medical Center
-
- Sanke Tokio
- Department of Clinical Laboratory Medicine, Wakayama Medical University
-
- Kuwashima Masamichi
- Department of Laboratory Medicine, University of Tokushima, School of Medicine
-
- Taminato Tomohiko
- Department of Laboratory Medicine, Kagawa University School of Medicine
-
- Ono Junko
- Department of Laboratory Medicine, Fukuoka University School of Medicine
Bibliographic Information
- Other Title
-
- ヘモグロビンA1C(IFCC値)およびグリコアルブミンの基準範囲の設定
- 委員会報告 ヘモグロビンA1C(IFCC値)およびグリコアルブミンの基凖範囲の設定
- イインカイ ホウコク ヘモグロビン A1C IFCCチ オヨビ グリコアルブミン ノ キジュン ハンイ ノ セッテイ
Search this article
Description
Since the hemoglobin A1C (HbA1C) measurement is to standardized internationally using references proposed by the International Federation of Clinical Chemistry and Laboratory Medicine (IFCC), we determined reference interval of HbA1C (IFCC). At the same time, we also determined reference interval of glycoalbumin (GA). The reference population (n=699 (men: 363, women: 336), age distribution: 23-91 years old) consisted of participants of the Funagata Diabetes Study and 3 other health checkup programs in Japan. All participants were subjected to a 75 g oral glucose tolerance test. HbA1C (IFCC) levels were measured with two types of routine HPLC and two types of routine immunoassay calibrated with JDS Lot 2 (assigned IFCC). Since no significant variation was found among the four measurements, the mean of these measurements was used as the representative value. GA levels in subjects were measured with an enzyme assay. Reference individuals for HbA1C (IFCC) and GA reference intervals were primarily selected based on National Committee of Clinical Laboratory Standards (NCCLS) recommendation. Multiple regression analysis showed that neither HbA1C (IFCC) or GA was associated with age, gender, questionnaire items suspected of influencing test results such as “Do you think you are healthy?,” physical examination results, or routine laboratory tests, e.g., hemoglobin, albumin, and creatinine. Reference individuals were selected from the reference population, with normal glucose tolerance, i.e., fasting plasma glucose<110 mg/dl and plasma glucose<140 mg/dl 2 hours after glucose loading. Reference individuals were further narrowed down using latent abnormal value exclusion method. Finally, using power transformation based on the parametric method, HbA1C (IFCC) and GA reference intervals were determined to be from 3.0 to 4.1%, and from 12.3 to 16.9%.
Journal
-
- Journal of the Japan Diabetes Society
-
Journal of the Japan Diabetes Society 49 (10), 825-833, 2006
THE JAPAN DIABETES SOCIETY
- Tweet
Keywords
Details 詳細情報について
-
- CRID
- 1390282679882464384
-
- NII Article ID
- 130004510933
- 10018393581
-
- NII Book ID
- AN00166576
-
- COI
- 1:CAS:528:DC%2BD2sXotVKksA%3D%3D
-
- ISSN
- 1881588X
- 0021437X
-
- NDL BIB ID
- 8581209
-
- Text Lang
- ja
-
- Data Source
-
- JaLC
- NDL Search
- CiNii Articles
-
- Abstract License Flag
- Disallowed