人工弁置換術後急性期ならびに慢性期における人工弁機能不全

DOI

書誌事項

タイトル別名
  • Acute and chronic dysfunction of the Bjork-Shiley type valve.

この論文をさがす

抄録

Two cases of acute valve dysfunction after mitral and aortic replacement with the Bjork-Shiley monostrut type valve were experienced. ECG and arterial pressure monitor and echo-cardiography were available in the diagnosis. This complication is not a usual one, but it occurs, an urgent re-operation must be performed as soon as possible And two cases of chronic valve disfunction after mitral replacement with the Bjork-Shiley convexo-concave type valve were experienced. They were fracture of the minor strut of the Bjork-Shiley valve. The only way of survival is immediate judgment of the eituatian and subsequent urgent operation for the valve dysfunction.