左房脱血補助人工心臓の限界
書誌事項
- タイトル別名
-
- Limitted functions of ventricular assist heart with an atrial withdrawal method.
この論文をさがす
説明
Pulsatile ventricular assist devices have successfully been in clinical use according to the Pierce's criteria. However, there are many cases with more critical cardiac functions. Atrial withdrawal cannula can not eliminate blood flow into ventricle. So ventricular myocardium can be minimized volume load but can not be reduced pressure after-loads. Subendocardial infarction was found at autopsy following transatiral VAD in the animal heart criated more critical myocardial disfunction.
収録刊行物
-
- 人工臓器
-
人工臓器 15 (2), 473-477, 1986