透析用flexible double lumen catheterの開発と臨床応用

DOI

書誌事項

タイトル別名
  • Flexible double-lumen catheter for temporary blood access.

この論文をさがす

抄録

Flexible double-lumen (FDL) catheter was designed as a temporary blond access (TBA). Silicone rubber is used for the material to enable the patients to walk around freely even if the FDL catheter is inserted through the femoral vein. An inner tube was placed inside of the lumen to insert the catheter using Seldinger's method. The maximum blood flow is more than 300ml/min and V pressure is below 100mmHg under the condition of 200ml/min. A new micropump using two syringes was designed to prevent clotting of the lumens by continuous in-jection of heparin. Clinical application was undertaken in 219 patients and the catheters were used for 1-4 weeks without serious complications FDL catheter is thought superior to external shunt because it can be used as a CVP or infusion line in addition to TBA.

収録刊行物

  • 人工臓器

    人工臓器 16 (2), 738-742, 1987

    一般社団法人 日本人工臓器学会