人工心臓の心拍出量制御による循環平衡特性の改善
書誌事項
- タイトル別名
-
- Improved circulatory autoregulation by driving pressure control of artificial heart
この論文をさがす
説明
The characterestics of the circulatory autoregulation during artificial heart pumping were studied, employing a simple mathematical model of the pulmonary and systemic circulation and the artificial heart. The unstable physiological state of arterial and venous pressure caused by the decreased cardiac output sensitivity to the venous pressure and the elevation of the pulmonary venous pressure more than 10 mmHg caused by the of terload effects, i.e., the decreases in cardiac output with increased arterial pressure, were predicted by numerical computations under the case of left, right or total artificial heart replacement. To improve the characterestics of the circulatory autoregulation, the servo-mechanism control and driving system of the total artificial heart was designed. In this system, cardiac output was regulated by the automatic changes in the driving pressure corresponding to the changes in the venous pressure, so that the venous pressure could be maintained at any constant level or within any desired range during the pumping.
収録刊行物
-
- 人工臓器
-
人工臓器 4 (6), 369-376, 1975
一般社団法人 日本人工臓器学会