Hi-Na dialysate parallel control system in central supply Method

DOI

Bibliographic Information

Other Title
  • セントラルサプライ式高Na透析液パラレルコントロール透析

Abstract

We have developed a controler system of sequential Hi/LO sodium concentration (144-134 mEq/1) in dialyzate for single pass method.<br>It was possible that in fact dialyzate sodium concentrations wave removed little by little and serum sodium slowly.<br>During benodialysis using this system, no trouble, such as dialysis disequilibrium syndrome was occured.<br>The merit of this system were the following:<br>(1) We don't need the I. V. infusion solution.<br>(2) The ultrafiltration rate was maintained above its desired level.<br>(3) No side-effect was found after dialysis such as hypertension.<br>(4) It is easy for us to treat this system in dialysate single pass method.