Investigation of Type I Thyroplasty Using Titanium Implant

Bibliographic Information

Other Title
  • チタン製インプラントを用いた甲状軟骨形成術Ⅰ型についての検討
  • チタンセイ インプラント オ モチイタ コウジョウ ナンコツ ケイセイジュツ Ⅰ カタ ニ ツイテ ノ ケントウ

Search this article

Description

Objective: The aim of this study was to describe a new titanium thyroplasty medializing implant and evaluate its usefulness.<br> Material and method: Retrospective study of 9 male patients with severe unilateral vocal fold paralysis after aortic-aneurysm surgery who underwent type I thyoplasty with arytenoid adduction under general anesthesia. Preoperative and postoperative glottal closure and voice function were evaluated by acoustic analysis (Jitter %, Shimmer %, NHR), stroboscopic findings, maximum phonation time, mean airflow rate, voice handicap index and computed tomography findings. These tests were evaluated before and 3 months after surgery.<br> Result: All acoustic parameters improved after surgery. In stroboscopic findings, symmetrical mucosal waves were identified in 3 cases. Maximum phonation time prolonged, mean flow rate decreased and voice handicap index improved. In computed-tomography findings, breakage, deformation, dropped and migration of titanium plate were not identified.<br> Conclusion: In surgery of vocal fold paralyzed patient, it is necessary to reconstruct vocal fold in physiological state at the time of phonation. Using the titanium plate developed for this purpose, all patients obtained good vocal improvement.

Journal

Citations (7)*help

See more

References(16)*help

See more

Details 詳細情報について

Report a problem

Back to top