内視鏡の汚染対策 気管支鏡の非定型抗酸菌汚染に対する簡易洗浄法について 特にブラッシングの有用性について
書誌事項
- タイトル別名
-
- A Simplified Cleaning Method of a Bronchoscope to Prevent Atypical Mycobacterial Contamination. Especially Concerning the Usefulness of Brushing.
- Especially Concerning the Usefulness of Brushing
- 特にブラツシングの有用性について
この論文をさがす
説明
The mycobacterial contamination of a bronchoscope has been a problem. At our institution, there was several cases of such contamination. Atypical mycobacterial contamination of tap water could not prevent by installation of 0.2μm filter system in the case of intermittent use of water. The highest contamination was found in the connection tube between filter systems. During the course of our trial to prevent atypical mycobacterial contamination, we recognized the importance of brushing. Using artificial sputum, we studied a simplified cleaning method for the bronchoscope. For the first step of cleaning a bronchoscope, we must physically wash it. For washing it, it is necessary to brush every part of the device 10 or more times. We test-produced a simplified cleaning device which uses 1.0kg/cm2 water pressure and benzalkonium chloride. It can be semi-automatically washed and it is useful because it is labor-saving. Because of its structure, suction valves were easily contaminated. Therefore, it needs to be carefully brushed, cleaned, and sterilized. The endoscope automatic washer (Olympus EW-10) is not perfect. The concentration of glutaraldehyde in EW-10 system decreased from 2 to 1.1% during one week. It is necessary to take care in having the user clean it by himself.
収録刊行物
-
- 日本気管食道科学会会報
-
日本気管食道科学会会報 43 (5), 395-402, 1992
特定非営利活動法人 日本気管食道科学会
- Tweet
詳細情報 詳細情報について
-
- CRID
- 1390282679990440576
-
- NII論文ID
- 130003669503
-
- ISSN
- 18806848
- 00290645
-
- 本文言語コード
- ja
-
- データソース種別
-
- JaLC
- Crossref
- CiNii Articles
- OpenAIRE
-
- 抄録ライセンスフラグ
- 使用不可