Studies on Blood Component Transfusion with Continuous Blood Cell Separator (IBM 2997)

Bibliographic Information

Other Title
  • 連続血液成分分離装置(IBM2997)による成分輸血の研究 I  成分血液採取の供血者に及ぼす影響について
  • 連続血液成分分離装置(IBM2997)による成分輸血の研究-1-成分血液採取の供血者に及ぼす影響について
  • レンゾク ケツエキ セイブン ブンリ ソウチ IBM2997 ニヨル セイブン
  • 第1編 成分血液採取の供血者に及ぼす影響について

Search this article

Description

We studied the safty of blood component extraction procedure with continuous-flow cell separator (IBM 2997) for normal donors. The changes of CBC and blood chemistry before and after blood component extraction were investigated.<br>1) The significant decrease of WBC, hemoglobin and hematocrit was seen in the leucocyte extraction group.<br>2) The significant decrease of platelets and WBC was seen in the platelet extraction group.<br>3) Indirect bilirubin, serum protein, albumin, calcium and inorganic phosphate decreased significantly after the procedure. Other liver and kidney function tests showed decreased values, because of dilution due to blood expansion by infusion of physiologic saline.<br>4) In three cases among 34 donors mild citrate adverse reaction were noted.<br>In conclusion, blood component extraction procedure with IBM 2997 is considerd a safe method for donors.

Journal

  • Rinsho Ketsueki

    Rinsho Ketsueki 22 (10), 1525-1530, 1981

    The Japanese Society of Hematology

Details 詳細情報について

Report a problem

Back to top