Accurate Measurement of Doses in the Sample Chamber of a <SUP>137</SUP>Csγ-ray Irradiation Apparatus
-
- ABE Toshiaki
- Radioisotope Research Center, University of Occupational and Environmental Health
-
- YAMAMOTO Hisao
- Radioisotope Research Center, University of Occupational and Environmental Health
-
- NORIMURA Toshiyuki
- Department of Radiation Biology and Health, School of Medicine, University of Occupational and Environmental Health
Bibliographic Information
- Other Title
-
- <SUP>137</SUP>Csガンマ線照射装置の試料容器内線量の精密測定
- 137Csガンマ線照射装置の試料容器内線量の精密測定
- 137Cs ガンマセン ショウシャ ソウチ ノ シリョウ ヨウキ ナイセンリョウ ノ セイミツ ソクテイ
Search this article
Abstract
Dose rates in the sample chamber of a 137Cs γ-ray irradiation apparatus (Nordion International Inc., Gammacell 40 Exactor) were accurately measured using an ionization chamber dosimeter. The dose rates obtained were 5.2m-7.2% higher at the upper and lower planes than at the middle plane in the sample chamber. In the process of starting and stopping irradiation, the additional dose during the movement of the source (end effects) was evaluated to be 0.85-1.21% (average 1.05±0.05%) of that estimated from the steady dose rate for 1min. When the irradiation period was less than 1 min, it was necessary to make a correction for the end effects in order to keep the error in irradiation dose below 1%.
Journal
-
- Journal of UOEH
-
Journal of UOEH 21 (4), 323-329, 1999
The University of Occupational and Environmental Health, Japan
- Tweet
Details 詳細情報について
-
- CRID
- 1390282680061276800
-
- NII Article ID
- 110001260003
-
- NII Book ID
- AN0009832X
-
- ISSN
- 21872864
- 0387821X
-
- NDL BIB ID
- 4914492
-
- Text Lang
- ja
-
- Data Source
-
- JaLC
- NDL
- Crossref
- CiNii Articles
-
- Abstract License Flag
- Disallowed