Investigation of Activation Rate Uniformity in a Prompt-.GAMMA. Rays IVNAA Facility

Bibliographic Information

Other Title
  • Investigation of activation rate uniformity in a prompt-γ rays IVNAA facility
  • Investigation of activation rate uniformity in a prompt g rays IVNAA facility

Search this article

Description

The preliminary design studies for developing a prompt-γ rays in vivo neutron activation analysis facility are in construction at Neutron Activation Research Centre of Ferdowsi University of Mashhad. In this work, activation rate distribution is studied in different bilateral configurations in order to reduce the CV of activation rate distribution throughout the body to less than 10%. It means that the maximum RMS deviation from the mean value of distribution must be smaller than 0.5%. The best case includes two pairs of 241Am-Be neutron sources. Each pair (with 70 cm distance between two sources) is positioned within up and down 40 × 110 cm2 area graphite collimators. Also, four moderator/reflector objects are added to transversal body surfaces with 5 cm thick water and 3 cm thick graphite. The proper thickness of polyethylene pre-moderators is chosen 8 mm. The RMS deviation of distribution is 0.3% with the mean activation rate of 1.85 × 10–6 cm–3.

Journal

References(35)*help

See more

Details 詳細情報について

Report a problem