Dosimetry Parameters of the IRH10 192Ir High Dose Rate Brachytherapy Source

  • YI Chul-Young
    Center for Ionizing Radiation, Korea Research Institute of Standards and Science
  • CHUN Kook Jin
    Center for Ionizing Radiation, Korea Research Institute of Standards and Science
  • HAH Suck-Ho
    Center for Ionizing Radiation, Korea Research Institute of Standards and Science
  • KIM Hyun-Moon
    Center for Ionizing Radiation, Korea Research Institute of Standards and Science
  • LIM Chunil
    Radiation Safety Division, National Institute of Food and Drug Safety Evaluation

この論文をさがす

抄録

The dosimetry parameters of the IRH10 192Ir high dose rate brachytherapy source were obtained from the dose calculation formalism recommended in the AAPM Task Group No. 43 report using the Monte Carlo code PENELOPE. The absorbed doses to water and air originating from the photons of the IRH10 192Ir brachytherapy source were calculated by the collision kerma approximation. The dose rate constant was evaluated to be (1.110 ± 0.011) cGy/h U–1. The dose rate per unit air kerma strength around the 192Ir IRH10 brachytherapy source and the anisotropy function were given in tables and figures.