Segmental Body Composition Assessment for Obese Japanese Adults by Single-Frequency Bioelectrical Impedance Analysis with 8-point Contact Electrodes

DOI 機関リポジトリ NDLデジタルコレクション NDLデジタルコレクション Web Site ほか1件をすべて表示 一部だけ表示 被引用文献4件 参考文献17件 オープンアクセス

書誌事項

公開日
2007
資源種別
journal article
DOI
  • 10.2114/jpa2.26.533
公開者
日本生理人類学会

この論文をさがす

説明

This study aimed to determine the accuracy of segmental body composition variables estimated by single-frequency BIA with 8-point contact electrodes (SF-BIA8), compared with dual-energy X-ray absorptiometry (DXA). Subjects were 72 obese Japanese adults (43 males and 29 females) aged 30 to 66 years. Segmental body composition variables (fat free mass: FFM, fat mass: FM, and percent fat mass: %FAT) were measured by these techniques. The correlations between impedance values and FFM measured by DXA were calculated. To examine the consistency in predicted values (SF-BIA8) with the reference (DXA), significant mean differences were tested by t-test and the degree of the difference was assessed by effect size. Correlations between the reference and predicted values were calculated. Additionally, the standard error of estimation (SEE) when estimating the reference from the predictor and the relative value of the SEE to the mean value of the DXA measurement (%SEE) were calculated. Systematic error was examined by Bland-Altman plots. High correlations were found between impedance and FFM measured by SF-BIA8. FFM in the extremities showed high correlations with the reference values, but systematic error was found. SF-BIA8 tended to overestimate FFM in the trunk. The consistencies in %FAT and FM with the reference value are inferior to those for FFM, and SEE values in %FAT and FM were greater than those for FFM. The accuracy of the estimated values in the trunk (FFM, %FAT, and FM) are inferior to those of the total body and extremities.

収録刊行物

被引用文献 (4)*注記

もっと見る

参考文献 (17)*注記

もっと見る

詳細情報 詳細情報について

問題の指摘

ページトップへ