- 【Updated on May 12, 2025】 Integration of CiNii Dissertations and CiNii Books into CiNii Research
- Trial version of CiNii Research Knowledge Graph Search feature is available on CiNii Labs
- 【Updated on June 30, 2025】Suspension and deletion of data provided by Nikkei BP
- Regarding the recording of “Research Data” and “Evidence Data”
-
- GOTO Kazuhiko
- Department of Science and Advanced Technology, Saga University
-
- SUGI Takenao
- Department of Advanced Technology Fusion, Saga University
-
- MATSUDA Yoshitaka
- Department of Advanced Technology Fusion, Saga University
-
- GOTO Satoru
- Department of Advanced Technology Fusion, Saga University
-
- FUKUDA Hiroki
- Yanagawa Rehabilitation School
-
- GOTO Yoshinobu
- Department of Occupational Therapy, International University of Health and Welfare
-
- YAMASAKI Takao
- Department of Clinical Neurophysiology, Kyushu University
-
- TOBIMATSU Shozo
- Department of Clinical Neurophysiology, Kyushu University
Bibliographic Information
- Other Title
-
- 視覚誘発電位記録状態の実時間評価システム
- シカク ユウハツ デンイ キロク ジョウタイ ノ ジツジカン ヒョウカ システム
Search this article
Description
Visual evoked potentials (VEPs) are widely used for the analysis and the diagnosis of the brain function on visual system. Since the accuracy of analysis depends on the quality of VEP record, the medical doctors and/or medical technologists always pay attention to the state of recording. In this study, the real-time evaluation system for VEP recording was newly developed. The system displayed the information for the characteristic parameters of VEPs, contaminated artifacts affecting VEP recording and the appearance of background EEG activity in real-time during the recording. Tendency of the VEP characteristics was also comparable with the past VEP records in the data base. Blink artifacts, electromyographic (EMG) artifacts and posterior alpha wave were detected for evaluating the state of subjects. Developed system was applied to VEP recordings of ten healthy adults, then the effectiveness of the proposed system was investigated. Results were revealed that the system achieved the quality of VEP recording with sufficient level and reduced the laborious and time consuming task during the recording. The developed system will be useful as an assistant tool in VEP recordings.
Journal
-
- Transactions of Japanese Society for Medical and Biological Engineering
-
Transactions of Japanese Society for Medical and Biological Engineering 50 (5), 433-442, 2012
Japanese Society for Medical and Biological Engineering
- Tweet
Details 詳細情報について
-
- CRID
- 1390282680243300224
-
- NII Article ID
- 130004494190
- 40019567027
-
- NII Book ID
- AA11633569
-
- ISSN
- 18814379
- 1347443X
-
- NDL BIB ID
- 024240591
-
- Text Lang
- ja
-
- Article Type
- journal article
-
- Data Source
-
- JaLC
- NDL Search
- CiNii Articles
- KAKEN
-
- Abstract License Flag
- Disallowed