- A New Method for Indirect Determination of Systolic and Diastolic Blood Pressure in the Radial Artery Using Inflection-Points of Pressure-Volume Curves Based on the Volume-Oscillometric Method
- 圧-容積曲線変曲点に着目した容積振動型収縮期・拡張期血圧決定法の提案
- アツ-ヨウセキ キョクセン ヘンキョクテン ニ チャクモク シタ ヨウセキ シンドウガタ シュウシュクキ ・ カクチョウキ ケツアツ ケッテイホウ ノ テイアン
この論文をさがす
抄録
A new method is presented for the indirect measurement of systolic (SBP) and diastolic blood pressure (DBP) in the radial artery. We use the volume-oscillometric method with a wrist cuff and then, utilizing the photo-plethysmograph (PPG) signal together with cuff pressure measurements, we derive pressure-volume curves (PVC). The principle of this method, named “PVC inflection-point method”, is to identify the indirect SBP and DBP using the inflection-points of envelope curves overlaying respectively the lower and upper points of the photo-plethysmographic pulsation signals superimposed on an arterial volume signal during the volume-oscillometric procedure. The accuracy of the method was evaluated by comparison with measurements from arterial catheterization in two healthy subjects. During the test procedure, a prototype PPG sensor for the PVC inflection-point method was attached to the left wrist, and arterial catheterization of the right radial artery for was performed to obtain simultaneous blood pressure recordings. The Bland-Altman plots for the comparative analysis showed good agreements between the two methods for both SBP (+2.3±7.7 (mean±S.D.) mmHg) and DBP (-1.8±6.4mmHg) measurement. We conclude that the proposed PVC inflection-point method appears promising to determine indirect SBP and DBP values with appropriate accuracy for convenient and practical use.
収録刊行物
-
- 生体医工学
-
生体医工学 49 (6), 968-976, 2011
公益社団法人 日本生体医工学会
- Tweet
キーワード
詳細情報 詳細情報について
-
- CRID
- 1390282680243624320
-
- NII論文ID
- 130004947519
-
- NII書誌ID
- AA12119469
-
- ISSN
- 18814379
- 1347443X
-
- NDL書誌ID
- 023569686
-
- 本文言語コード
- ja
-
- データソース種別
-
- JaLC
- IRDB
- NDL
- CiNii Articles
- KAKEN
-
- 抄録ライセンスフラグ
- 使用不可