- Correction for X-ray Spectra Measured with a CdZnTe Detector (2)
この論文をさがす
説明
X-ray spectra measured with a CdZnTe detector were corrected using response functions calculated with the Monte Carlo method. In the calculation, the Hecht equation was employed to deal with the effects of carrier trapping in a CdZnTe crystal. Corrected spectra showed excellent agreement with x-ray spectra obtained with a HPGe detector. This meant that the response functions calculated in this study were valid. We conclude that a CdZnTe detector can be employed in diagnostic x-ray spectroscopy with proper corrections.