Development of Parallel Imaging Program
-
- IMAI Hiroshi
- Division of Health Sciences, Graduate School of Medical Science, Kanazawa University
-
- KASUGA Toshio
- Department of Radiological Technology, School of Health Sciences, Faculty of Medicine, Kanazawa University
-
- MIYATI Toshiaki
- Department of Radiological Technology, School of Health Sciences, Faculty of Medicine, Kanazawa University
-
- KASAI Masaharu
- Department of Central Radiology, Nagoya City University Hospital
-
- KAWANO Makoto
- Department of Central Radiology, Nagoya City University Hospital
-
- SUGIYAMA Masayuki
- Department of Central Radiology, Nagoya City University Hospital
Bibliographic Information
- Other Title
-
- パラレルイメージングプログラムの開発
Description
Parallel imaging in magnetic resonance imaging (MRI) is a method to reduce the acquisition time by using phased array coils. However, there are a small number of commercial diagnostic MR imagers which could be used that method. Therefore, we developed a parallel imaging program for sensitivity encoding (SENSE). We produced sensitivity maps calculated from phantom images which was obtained with each coil element. Next, we eliminated wrap-around artifacts by using the sensitivity maps. Parallel imaging with SENSE made it possible to achieve fast scan or remove wrap-around artifacts.
Journal
-
- Medical Imaging and Information Sciences
-
Medical Imaging and Information Sciences 20 (1), 27-33, 2003
Medical Imaging and Information Sciences
- Tweet
Details 詳細情報について
-
- CRID
- 1390282680431967488
-
- NII Article ID
- 130004647664
-
- ISSN
- 09101543
-
- Text Lang
- ja
-
- Data Source
-
- JaLC
- CiNii Articles
-
- Abstract License Flag
- Disallowed