書誌事項
- タイトル別名
-
- Correlation between Required Quality Assurance Equipment and Radiation Therapy Institutional Level
- ホウシャセン チリョウ シセツ レベル ト セイド カンリ ニ ヒツヨウ ナ キキ
この論文をさがす
抄録
Radiotherapy institutions carry out various types of treatments such as conventional radiation treatment, stereotactic radiotherapy (SRT), intensity-modulated radiation therapy (IMRT) and image-guided radiotherapy (IGRT). According to the radiation treatment type, various kinds of quality assurance (QA) equipment are used to maintain the accurate dose delivery to a patient with accurate positioning. The aim of this report was to propose the QA equipment required at various institutional levels. The institution levels were categorized according to the radiation treatment cases. Institutions in the basic category used conventional radiotherapy techniques employing field sizes larger than 4 cm×4 cm, or they had no full-time radiation oncologist. In addition to the basic level, five other categories were defined which were based on being able to carry out total body irradiation, three-dimensional conformal radiation therapy / SRT, IMRT, and IGRT. Minimum QA equipment requirements were proposed for each level. This proposal will facilitate the purchase of QA equipment for the installation or refurbishing of a linear accelerator.
収録刊行物
-
- 医学物理
-
医学物理 31 (3), 84-89, 2011
公益社団法人 日本医学物理学会
- Tweet
詳細情報 詳細情報について
-
- CRID
- 1390282680487252992
-
- NII論文ID
- 130005117995
-
- NII書誌ID
- AA11580542
-
- ISSN
- 21869634
- 13455354
-
- NDL書誌ID
- 023660369
-
- 本文言語コード
- ja
-
- データソース種別
-
- JaLC
- NDL
- CiNii Articles
-
- 抄録ライセンスフラグ
- 使用不可