Measurement of trace light hydrocarbons in methane-rich gases

Bibliographic Information

Other Title
  • メタンを主成分とするガス中の微量軽質炭化水素の測定法
  • メタン オ シュセイブン ト スル ガスチュウ ノ ビリョウ ケイシツ タンカ

Search this article

Description

A system to measure C2+ light hydrocarbons (CnHm, n≧2) in methane-rich gases is reported. Until now, measurement of trace C2+ light hydrocarbons by gas chromatography (GC) has been difficult, because C2+ hydrocarbons could not be separated well from methane. In this system, C2+ hydrocarbon concentration column (stainless hollow pipe) is equipped between GC injection and GC column. With the C2+ hydrocarbon concentration column, only C2+ hydrocarbons are concentrated and methane is removed. After the removal of methane, precise measurement of C2+ hydrocarbons by GC becomes possible.

Journal

  • Chikyukagaku

    Chikyukagaku 29 (1), 17-23, 1995

    The Geochemical Society of Japan

Citations (2)*help

See more

References(9)*help

See more