Effects of concentration of Na<sub>2</sub>CO<sub>3</sub>-containing stabilizer on urinary C-peptide examination
-
- MURAKAMI Mariko
- Department of Clinical Laboratory, National Cerebral and Cardiovascular Center
-
- ASADA Takashi
- Department of Clinical Laboratory, National Cerebral and Cardiovascular Center
-
- OZAWA Kazami
- Department of Clinical Laboratory, National Cerebral and Cardiovascular Center
-
- NOGAMI Tsuyoshi
- Department of Clinical Laboratory, National Cerebral and Cardiovascular Center
-
- HATSUDA Kazuyoshi
- Department of Clinical Laboratory, National Cerebral and Cardiovascular Center
-
- SANO Michitaka
- Department of Clinical Laboratory, National Cerebral and Cardiovascular Center
Bibliographic Information
- Other Title
-
- 尿中Cペプチド検査における安定化剤含有濃度の影響
Search this article
Description
Urinary C-peptide examination is used as an indirect method of assessing the level of insulin production. We use 24-h urine samples as materials for examination. We use a Na2CO3-containing stabilizer of urinary C-peptide. Under acidic conditions, C-peptide is unstable and is decomposed by growing bacteria. However, the effect of the concentration of the Na2CO3-containing stabilizer on the measurements of collected urine volume is unknown. We examined the effects of the concentration of the Na2CO3-containing stabilizer, storage temperature, and bacterial presence on the measurements of collected urine volume. According to our results, CPR was decomposed by growing bacteria, and the urinary CPR concentration decreased when the collected urine volume per package of the stabilizer exceeded 4,000 mL at room temperature. When we chilled and preserved urine with low concentrations of the stabilizer, CPR was stable. In addition, in the case of the collected urine volume that was less than 300 mL per package of the stabilizer, the urine CPR concentration decreased. After examining the effects of both pH and salt concentration, we considered that salt concentration has a greater effect than pH on CPR when a high concentration of the stabilizer is used. Therefore, we must determine the appropriate concentration of the Na2CO3-containing stabilizer for a particular collected urine volume before measurements.
Journal
-
- Japanese Journal of Medical Technology
-
Japanese Journal of Medical Technology 64 (5), 576-582, 2015
Japanese Association of Medical Technologists
- Tweet
Details 詳細情報について
-
- CRID
- 1390282680717730944
-
- NII Article ID
- 130005109640
-
- ISSN
- 21885346
- 09158669
-
- Text Lang
- ja
-
- Data Source
-
- JaLC
- CiNii Articles
-
- Abstract License Flag
- Disallowed