Novel Cardiac SPECT Technology with Semiconductor Detectors : Emerging Trends and Future Perspective
-
- Miyagawa Masao
- Department of Radiology, Ehime University Graduate School of Medicine
-
- Nishiyama Yoshiko
- Department of Radiology, Ehime University Graduate School of Medicine
-
- Tashiro Rami
- Department of Radiology, Ehime University Graduate School of Medicine
-
- Ishimura Hayato
- Department of Radiology, Ehime University Graduate School of Medicine
-
- Takahashi Yasuyuki
- Department of Nuclear Medicine Technology, Gunma Prefectural College of Health Sciences
-
- Mochizuki Teruhito
- Department of Radiology, Ehime University Graduate School of Medicine
書誌事項
- タイトル別名
-
- Novel Cardiac SPECT Technology with Semiconductor Detectors
- Emerging Trends and Future Perspective
この論文をさがす
抄録
Recently, two vendors have introduced novel SPECT scanners with solid-state semiconductor detectors: Discovery NM530c (D530c) and D-SPECT utilizing the same cadmium zinc-telluride (CZT) detectors, with a different combination of high-sensitivity multi-pinhole or parallel-hole collimator which focuses on the myocardium. The physical performance is dramatically higher than that of conventional Anger cameras, however, 2 CZT cameras are inherently different. Spatial resolution and contrast-to-noise ratio are better with the D530c, whereas detection sensitivity is higher with the D-SPECT. Although 99mTc-labeled myocardial perfusion tracers might not be ideal, estimation of absolute myocardial blood flow or coronary flow reserve using dynamic CZT SPECT is of great interest and a challenging subject.
収録刊行物
-
- Annals of Nuclear Cardiology
-
Annals of Nuclear Cardiology 1 (1), 18-26, 2015
日本心臓核医学会