S0220203 Development of phantom experiment method with flexible stenosis tube for fluid simulation

  • GORYU Akihiro
    Corporate Research and Development Center, Toshiba Corporation
  • HONGO Takuya
    Corporate Research and Development Center, Toshiba Corporation
  • KANO Akira
    Corporate Research and Development Center, Toshiba Corporation
  • KATO Mitsuaki
    Corporate Research and Development Center, Toshiba Corporation
  • HIGASHI Shinya
    Corporate Research and Development Center, Toshiba Corporation
  • HIROHATA Kenji
    Corporate Research and Development Center, Toshiba Corporation

Bibliographic Information

Other Title
  • S0220203 狭窄を有する柔軟管による流体解析用ファントム実験手法の開発

Description

In this paper, we described a novel phantom experiment method with flexible stenosis tube, for use of circulatory simulation by fluid-structure coupled analysis. Recently, factional flow reserve (FFR) for evaluation approach of coronary stenosis has received a lot of attention. However previous technics is still difficult to measure the FFR values, because catheter surgery is required. To solve this problem, we have developed a novel approach based on four-dimensional computed tomography (4D-CT) image tracking and structural and fluid analysis named 4D-CT FFR, promising minimally invasive for patients. Here we demonstrated phantom experiment with flexible stenosis tube, and indicated relationship between deformations of the tube, pressure loss and flow rate, for validation of 4D-CT-FFR.

Journal

Citations (1)*help

See more

Details 詳細情報について

Report a problem

Back to top