抄録
Particle Image Velocimetry (PIV) analysis was conducted for a newly developed mass flow meter especially for axial flow blood pumps. As a result, it was found that the flow patterns are different between low flow and high flow conditions depending on the flow rate. This result explains the difference of calibration coefficients between over and below the flow of 1.0 L/min.