413 PIVによる曲がり管式質量流量計内の速度分布の計測(OS5-(4)オーガナイズドセッション《ライフサポートにおける工学技術》)

書誌事項

タイトル別名
  • 413 Measurement of Velocity Distribution by PIV in a Newly Developed Mass Flow Meter

抄録

Particle Image Velocimetry (PIV) analysis was conducted for a newly developed mass flow meter especially for axial flow blood pumps. As a result, it was found that the flow patterns are different between low flow and high flow conditions depending on the flow rate. This result explains the difference of calibration coefficients between over and below the flow of 1.0 L/min.