516 生体組織融着技術を用いたステントの位置ずれ防止:小型コイルによるステントの誘導加熱と引張強度評価(OS5 オーガナイズドセッション《ライフサポートにおける工学技術》)

書誌事項

タイトル別名
  • 516 The Prevention of Stent Migration Using Tissue Adhesion Method:lnductive Heating of The Stent with Small Coil and Evaluation of Tensile Strength

説明

We propose a technique to prevent the migration using the biomedical tissue adhesion technology. In this paper, heat characteristic of stent and adhesive strength between stent and tissues were measured. The tissues were porcine arteiy and esophagus. Frequency, input current, and measurement time were 600 kHz, 3.0 A, and 180 s, respectively. As a result, the time that porcine artery and esophagus temperature reached at 80 ℃ were 136 s and 68 s, respectively. 0.02 MPa can be obtained as adhesive strength for both of them.