書誌事項
- タイトル別名
-
- Mechanically Separable Master Arm Possible to be Autoclave Sterilized for Surgical Support Robot
抄録
<p>A new mechanically separable master device has developed that can be autoclave sterilized to master-slave control for a surgery support forceps manipulator (LODEM) with five-DOFs. The master device is separated into the handy mechanical part and the encoder sensor part connected to driving wires. The mechanical part uses a ball joint and gimbals for pitch and yaw axes, and linear guide for insertion axis. The operating range is ±70° for the pitch and yaw axes, and 0 to 100 mm for the insertion/extraction axis. The dimensions and the mass of the mechanical part are 126*126*345 mm and 1.0 kg. The dimensions and the mass of the sensor part are 160*87*200 mm and 1.4 kg. The backlash was 0.8°, 1.8°, and 0.6 mm respectively.</p>
収録刊行物
-
- ロボティクス・メカトロニクス講演会講演概要集
-
ロボティクス・メカトロニクス講演会講演概要集 2017 (0), 1P1-K05-, 2017
一般社団法人 日本機械学会
- Tweet
詳細情報 詳細情報について
-
- CRID
- 1390282680917411968
-
- NII論文ID
- 130006220524
-
- ISSN
- 24243124
-
- 本文言語コード
- ja
-
- データソース種別
-
- JaLC
- Crossref
- CiNii Articles
-
- 抄録ライセンスフラグ
- 使用不可