- 安田 光慶
- Faculty of Health Sciences, Komazawa University Department of Radiological Technology, Showa University Fujigaoka Hospital
-
- 加藤 京一
- Graduate School of Nursing and Rehabilitation Sciences, Showa University
-
- 黒澤 駿
- Department of Radiological Technology, Showa University Fujigaoka Hospital
-
- 吉川 宏起
- Faculty of Health Sciences, Komazawa University
-
- 佐藤 久弥
- Graduate School of Nursing and Rehabilitation Sciences, Showa University
-
- 高橋 俊行
- Graduate School of Nursing and Rehabilitation Sciences, Showa University
-
- 岩井 譜憲
- Graduate School of Nursing and Rehabilitation Sciences, Showa University
-
- 渡辺 裕之
- Graduate School of Nursing and Rehabilitation Sciences, Showa University
-
- 中澤 靖夫
- Graduate School of Nursing and Rehabilitation Sciences, Showa University
書誌事項
- タイトル別名
-
- Observations of Deterioration of Photo-Stimulable Phosphor Plates Used in Digital Mammography and Its Effects on Mammography Diagnoses
- ディジタルマンモグラフィ ノ キジンセイ ケイコウタイ プレート レッカ ニ ヨル リンショウ ガゾウ エ ノ エイキョウ
この論文をさがす
説明
We noted that breast-like artifacts occur in photo-stimulable phosphor plates (PSPs) used for long periods in digital mammography systems. This prompted us to investigate the effects on mammography diagnoses of these artifacts. Our study took the form of a comparison between the images generated by a long-term use PSP (four years use) and a short-term use PSP. First, an acrylic phantom and an in-house-made phantom for visual evaluation were imaged using X-ray PSPs under the same exposure conditions. They were then scanned under the same conditions to generate images. The mean values of the digital signals were measured in the 35 locations of region of interest present in the images of the acrylic phantom. The images of the in-house-made phantom were then visually evaluated by five certified experts, who examined 22 clinical images taken with both PSPs while referring to a visual evaluation scale. Differences were detected in all the evaluated items: the short-term use PSP gave a higher rate than the long-term use PSP (p<0.01), suggesting that long-term use PSPs might adversely affect the results of diagnostic mammography.
収録刊行物
-
- 日本放射線技術学会雑誌
-
日本放射線技術学会雑誌 69 (4), 393-399, 2013
公益社団法人 日本放射線技術学会
- Tweet
詳細情報 詳細情報について
-
- CRID
- 1390282681337764096
-
- NII論文ID
- 10031159461
-
- NII書誌ID
- AN00197784
-
- COI
- 1:STN:280:DC%2BC3srns1ajtw%3D%3D
-
- ISSN
- 18814883
- 03694305
-
- NDL書誌ID
- 024691347
-
- PubMed
- 23609861
-
- 本文言語コード
- ja
-
- データソース種別
-
- JaLC
- NDLサーチ
- Crossref
- PubMed
- CiNii Articles
- OpenAIRE
-
- 抄録ライセンスフラグ
- 使用不可