ON THE SCATTERING DISTRIBUTION OF VARIOUS X-RAY APPARATUS : 2nd REPORT FLUOROSCOPY APPARATUS
-
- NAKAMURA M.
- Dipt. Radiorogy Shiohama Hospital Mie University
-
- KAWAMOTO S.
- Dipt. Radiorogy Shiohama Hospital Mie University
-
- KUSUNOSE M.
- Dipt. Radiorogy Shiohama Hospital Mie University
Bibliographic Information
- Other Title
-
- 各種X線装置の散乱線分布状態に就いて : 第2報透視診断用裝置
- 各種X線装置の散乱線分布状態に就いて-2-
- カクシュ Xセン ソウチ ノ サンラン センブンフ ジョウタイ ニ ツイテ 2
Search this article
Abstract
We mcasured the scattered radiation around the various kinds of flunroscopy tables with Pocketchambers or Ionizationchambers survey meter. A Phantom or a Patient's body was the some of the s.r. of which the back s.r. was the softs and strongest. To protec〓 the back s.r. we studied the effect of a aluminium plate placed under the table and coneluded that a Al plate of 1mm thickness could reduce the back s.r. to about 50%. A new type of tables, which had the iron plate and lead gummy shcets around the patient's body, was proved to be completely safe as regard to the present softy standard.