最新の80対マルチリーフコリメータの物理特性

書誌事項

タイトル別名
  • Physical Characterizations for an Integrated 160-leaf Multi-leaf Collimator with a New Concept Design
  • サイシン ノ 80ツイ マルチリーフコリメータ ノ ブツリ トクセイ

この論文をさがす

説明

In this article, we present a physical characterization of the agility (Elekta). agility is composed of 160 interdigitating multileaf collimators (MLCs) with a width of 5 mm at the isocenter. The physical characterizations that include leaf position accuracy, leakage, field penumbra and the tongue-and-groove (T&G) effect were evaluated using well-commissioned 4, 6 and 10-MV photon beams. The leaf position accuracy was within 0.5 mm for all gantry angles and each MLC. The leakage was 0.44% on average and reached 0.47% at 10 MV: remarkably low due to a new design with tilted leaves. However, the T&G effect occurred due to tilt. It was approximately 20.8% on average and reached 22.3% at 6 MV. The penumbra width increased up to 8.5 mm at a field size of 20×20 cm at 4 MV. High position designed MLCs create a wider penumbra but show lower leakage and large head clearance. Head clearance is an important factor in stereotactic radiotherapy with multiple non-coplanar beams.