A System of Setting Exposure Conditions in General X-rays Using a Calculating Formula
-
- Ogawa Kenichi
- Kyoto University Hospital, Clinical Radiology Service
Bibliographic Information
- Other Title
-
- 一般X線撮影時の照射条件設定システムの構築
- 臨床技術 一般X線撮影時の照射条件設定システムの構築
- リンショウ ギジュツ イッパン Xセン サツエイジ ノ ショウシャ ジョウケン セッテイ システム ノ コウチク
Search this article
Description
The purpose of this study was to determine the exposure conditions for general X-rays by calculation. The concept of the basic equation for the calculation is that the dose after transmitting an object is determined by entering the exposure conditions. For this purpose, we formulated a basic equation by mathematically representing all factors that influence imaging conditions. Using the basic equation, we succeeded in determining the optimal exposure condition for a patient’s body thickness by calculation. Even if another equipment, grid, added filter, etc., is used, optimal exposure conditions can be determined by adding correction factors determined by simple added measurements; this confirmed the versatility of the system. The calculation accuracy was verified to be within 15%. By using this system, we succeeded in optimizing exposure conditions for general X-rays.
Journal
-
- Japanese Journal of Radiological Technology
-
Japanese Journal of Radiological Technology 66 (8), 885-892, 2010
Japanese Society of Radiological Technology
- Tweet
Details 詳細情報について
-
- CRID
- 1390282681338892672
-
- NII Article ID
- 10026567858
-
- NII Book ID
- AN00197784
-
- ISSN
- 18814883
- 03694305
-
- NDL BIB ID
- 10818485
-
- PubMed
- 20953104
-
- Text Lang
- ja
-
- Data Source
-
- JaLC
- NDL Search
- Crossref
- CiNii Articles
- OpenAIRE
-
- Abstract License Flag
- Disallowed