書誌事項
- タイトル別名
-
- Effect of Reconstruction Technique for Metal Artifact Reduction in Computed Tomography by Changing Display Field of View
- 臨床技術 金属アーチファクト低減再構成技術におけるdisplay field of view(DFOV)の影響について
- リンショウ ギジュツ キンゾク アーチファクト テイゲン サイコウセイ ギジュツ ニ オケル display field of view(DFOV)ノ エイキョウ ニ ツイテ
この論文をさがす
説明
We evaluated the effect of orthopedic-metal artifact reduction (O-MAR) for metal artifact in computed tomography with 73 simulated seeds for brachytherapy in different sizes of display field of view (DFOV) obtained by helical scan under the same clinical scan condition. The metal artifacts were analyzed with the Gumbel's method by changing DFOV sizes 80 mm, 160 mm, and 320 mm. Gumbel distribution, scale parameter (γ), and location parameter (β) of the metal artifacts with O-MARwere compared with that of the metal artifacts with filtered back projection (FBP). In conclusion, it was considered that the effect of metal artifact reduction with O-MARwas influenced by DFOV size in this study. The reduction rates of scale parameter (γ) were 22.3%, 21.3%, and 10.0% in DFOV 80 mm, 160 mm, and 320 mm, respectively. The reduction rates of location parameter (β) were 27.4%, 23.4 %, and 9.8%. Therefore, the effect of metal artifact reduction with O-MARshowed the tendency of increasing with decreasing DFOV size.
収録刊行物
-
- 日本放射線技術学会雑誌
-
日本放射線技術学会雑誌 71 (11), 1096-1102, 2015
公益社団法人 日本放射線技術学会
- Tweet
キーワード
詳細情報 詳細情報について
-
- CRID
- 1390282681339243904
-
- NII論文ID
- 130005109945
-
- NII書誌ID
- AN00197784
-
- ISSN
- 18814883
- 03694305
-
- NDL書誌ID
- 026954491
-
- PubMed
- 26596201
-
- 本文言語コード
- ja
-
- データソース種別
-
- JaLC
- NDL
- Crossref
- PubMed
- CiNii Articles
-
- 抄録ライセンスフラグ
- 使用不可