- A Novel Evaluation Method for Displacement between Carbon Beam Axis and Positioning X-ray Axis Using an Imaging Plate
- Imaging Plate オ モチイタ タンソセン ト イチ ショウゴウヨウ Xセンジク ノ ヘンイリョウ ヒョウカ ホウホウ ノ カイハツ
この論文をさがす
抄録
Purpose: We developed an evaluation method for easily calculating displacement directly between the carbon beam axis and positioning X-ray axis. Methods: A verification image was acquired by irradiating an imaging plate with a carbon beam and X-ray. The X-ray passed through a lead plate inserted in the range compensator holder. The displacement was calculated on the verification image from the center of a wire irradiated with carbon using a multi leaf collimator (MLC) and a wire irradiated with X-ray also using MLC. The accuracy of the method was evaluated by moving the carbon beam axis, the X-ray axis, and the setup angle. The weekly changes of vertical and lateral beams in all rooms were also evaluated. Results: The displacements of the carbon beam axis and the setup angle did not influence the calculation results, whereas the displacement of the X-ray axis did (R=0.999). The displacements including weekly changes were all less than 1.00 mm. Conclusion: An evaluation method for calculating the displacement directly and simply between the carbon beam axis and positioning X-ray axis was developed and verified. The weekly changes of displacement between axes were evaluated to be acceptable at our facility.
収録刊行物
-
- 日本放射線技術学会雑誌
-
日本放射線技術学会雑誌 72 (2), 139-148, 2016
公益社団法人 日本放射線技術学会
- Tweet
キーワード
詳細情報 詳細情報について
-
- CRID
- 1390282681339370880
-
- NII論文ID
- 130005126330
-
- NII書誌ID
- AN00197784
-
- ISSN
- 18814883
- 03694305
-
- NDL書誌ID
- 027186288
-
- PubMed
- 26902378
-
- 本文言語コード
- ja
-
- データソース種別
-
- JaLC
- NDL
- Crossref
- PubMed
- CiNii Articles
-
- 抄録ライセンスフラグ
- 使用不可