Overview of the Infant Body Part Photography Condition Using 16-rows Multidetector CT

  • Futamata Yoshihiro
    JA Hokkaidoukouseiren Obihiro Kousei Hospital, Medical Technology Section, Radiation Technological Department
  • Takahashi Yasutomo
    JA Hokkaidoukouseiren Obihiro Kousei Hospital, Medical Technology Section, Radiation Technological Department
  • Kikuchi, Takahiro
    JA Hokkaidoukouseiren Obihiro Kousei Hospital, Medical Technology Section, Radiation Technological Department
  • Sugiyama Sunao
    JA Hokkaidoukouseiren Obihiro Kousei Hospital, Medical Technology Section, Radiation Technological Department

Bibliographic Information

Other Title
  • 16列MDCTにおける小児体幹部撮影条件の基礎的検討
  • 臨床技術 16列MDCTにおける小児体幹部撮影条件の基礎的検討
  • リンショウ ギジュツ 16レツ MDCT ニ オケル ショウニ タイカンブ サツエイ ジョウケン ノ キソテキ ケントウ

Search this article

Description

In infant CT scans, it is important to minimize radiation exposure without lowering the quality of the diagnosis. Therefore, appropriate parameters for infant CT scan should be considered at each institute. In order to determine parameters for infant body CT, we measured the physical characteristics of our current CT machine and evaluated scan parameters as S.D. below 10 on the basis of literature recommendations, which say that we should adopt S.D. of the liver when treated with a radiation exposure dose in an adult abdominal CT scan. As a result, the ideal parameters were 90 kV tube voltage, ultra fast detail resolution, B kernel, and Eff.mAs value calculated from patient body width. Infant body CT scans with the mentioned parameters resulted in an S.D. below 10, and this is thought to be applicable to further examinations.

Journal

References(12)*help

See more

Details 詳細情報について

Report a problem

Back to top