- 光田 秀雄
- 大阪市立大学医学部放射線医学教室
書誌事項
- タイトル別名
-
- STUDY ON ERROR IN MEASUREMENT WITH FLUORO GLASS DOSIMETER
- ケイコウ ガラスセンリョウケイ ニ ヨル ソクテイ ニ オケル ゴサ ノ ケントウ
この論文をさがす
抄録
The author studied an error of numerical value in measurement with fluoro glass dosimeter. The fluoro glass dosimeter. The fluoro glass used in this study was FD-R1-1 (Toshiba) and had a diameter of 1mm and a length of 6mm. The fluorh meter was FGD-3B (Toshiba). The author applied electron beam of Betatron to the fluoro glass dosimeter in a phantom made of acrylite. The fluctuation of the numerical value of the fluoro meter in a short time was about ±0.15R ; the error of the numerical value, which was probably due to the play of the apparatus, was about 1.56%. The response of the fluoro glass showed a good linearity, extending 1-10^3R of the fluorescence ; the fluorescence of the back ground of the fluoro glass was about 0.7±0.07R (O/x=10%). In this experiment, the author studied also the response of the fluoro glass, using about 100 fluoro glasses. The result was as follows ; Fluorescence The standard deviation in the response [table] It may be said through the comparative study over foreign literature that these numerical value showed better efficiency of this fluoro glass dosimeter the that of foreign ones. The fading of the fluorescence reaches to 5% in 100days and the value of the fading is apot to be divided. The fading is caused also by repeating the measurement of the fluorescence. The author made a futher examination on the error in the measurement with fluoro glass dosimeter through these results mentioned above and other literatures.
収録刊行物
-
- 日本放射線技術学会雑誌
-
日本放射線技術学会雑誌 23 (2), 202-211, 1967
公益社団法人 日本放射線技術学会
- Tweet
キーワード
詳細情報 詳細情報について
-
- CRID
- 1390282681341799808
-
- NII論文ID
- 110003453095
-
- NII書誌ID
- AN00197784
-
- ISSN
- 18814883
- 03694305
-
- NDL書誌ID
- 8403576
-
- 本文言語コード
- ja
-
- データソース種別
-
- JaLC
- NDL
- Crossref
- CiNii Articles
-
- 抄録ライセンスフラグ
- 使用不可