- 稲津 博
- 宮崎県立宮崎病院放射線科
書誌事項
- タイトル別名
-
- SPATIAL FREQUENCY CHARACTERISTICS OF REVERSAL RADIOGRAPH
- 反転X線写真の空間習波数特性
- ハンテン Xセン シャシン ノ クウカン シュウハスウ トクセイ
この論文をさがす
説明
To improve the film graininess, we have tried the reversal development method. As the larger silver grains, which are more sensitive than smaller ones, are removed more than smaller ones from the emulsion of film in the process of bleaching, the graininess of the reversally developed film is much better than that of the original negative film. Film graininess is improved especially at the part of higher density of negative X-ray film. We investigated the modulation transfer function, autocorrelation function and Wiener spectra as to reversal and negative films. The results show that MTF of reversal and negative X-ray films are almost same and so depend only on the sensitivity of intensifying screens.
収録刊行物
-
- 日本放射線技術学会雑誌
-
日本放射線技術学会雑誌 31 (5), 426-430, 1976
公益社団法人 日本放射線技術学会