RELATIONSHIP BETWEEN DETECTABILITY BY THE 18-AFC METHOD AND SPECTRAL SN RATIO
-
- OKADA HIROHARU
- Department of Central Radiology, Kinki University Hospital
-
- USAMI KIMIO
- Department of Central Radiology, Kinki University Hospital
-
- FUKUI YUTAKA
- Department of Central Radiology, Kinki University Hospital
-
- NISHIMATSU KAZUHIKO
- Department of Central Radiology, Kinki University Hospital
-
- HASHIBA HISAYUKI
- Department of Central Radiology, Kinki University Hospital
-
- NANBU HIDEKAZU
- Department of Central Radiology, Kinki University Hospital
-
- TATSUMI SACHIYO
- Department of Central Radiology, Kinki University Hospital
-
- SAKASHITA TARO
- Department of Central Radiology, Kinki University Hospital
-
- MURANO YOSHIHIKO
- Department of Central Radiology, Kinki University Hospital
-
- WAKAMATSU TAKASHI
- Department of Radiology, Utano National Hospital
-
- YAMASHITA KAZUYA
- College of Bio-Medical Technology, Osaka University
Bibliographic Information
- Other Title
-
- 18 肢強制選択法の検出率とスペクトル SN 比との相関について
Search this article
Description
We examined the relationship between SN ratio of visual threshold value and radiographic image by using various sizes of plastic columns and various speeds of screen/film systems to change the signal value. We tried to get a visual estimation applyjng the 18-AFC (18-alternative forced choice) method using five radiological technologists. On the other side, we got a physical estimation applying the spectral SN ratio composed of NEQ (u) (Number of noise-equivalent quanta (u)) and distributed spectral of the signal. As a result, when we assumed the threshold value of detectability of the 18-AFC method was 50%, we found that SN ratio of the visual threshold value was 3.23. When the speed of the screen/film system increases, the threshold value of the SN ratio decreases. We were able to confirm that the visual characteristics depend on the contrast and spatial frequency of the signal.
Journal
-
- Japanese Journal of Radiological Technology
-
Japanese Journal of Radiological Technology 49 (5), 682-690, 1993
Japanese Society of Radiological Technology
- Tweet
Keywords
Details 詳細情報について
-
- CRID
- 1390282681356082304
-
- NII Article ID
- 110003471105
-
- ISSN
- 18814883
- 03694305
-
- Text Lang
- ja
-
- Data Source
-
- JaLC
- Crossref
- CiNii Articles
- OpenAIRE
-
- Abstract License Flag
- Disallowed