X線CTにおけるX線管回転速度の高速化が画質に与える影響

  • 山口 功
    国家公務員共済組合連合会呉共済病院放射線部:(現)福井大学医学部附属病院放射線部
  • 森本 章
    国家公務員共済組合連合会呉共済病院放射線部

書誌事項

タイトル別名
  • Influence Rotation Speed of X-ray Computed Tomography on Image Quality

この論文をさがす

説明

The rotation speed of X-ray CT is known to influence the quality of images. We examined changes in imaging characteristics according to various physical characteristics. The physical characteristic of rotation speed (1.0, 0.8, 0.7, 0.6, 0.5 sec/rot) was evaluated at the FOV center (isocenter) and 90 mm outside the center (off-center). The items of evaluation included low contrast resolution, high contrast resolution, noise, definition, and artifacts. High contrast resolution and definition declined at a rotation speed of O.6 sec/rot or less at off-center. An increase in shower artifact and change in the shape of the wire section were found at a rotation speed of 0.6 sec/rot or less at off-center. It was determined that image quality was especially influenced in terms of definition characteristics when the speed of rotation was high. A decrease in the number of views can be considered the main cause ; that is, the use of high-speed rotation degrades image quality. Therefore, it is important to consider proper usage as it corresponds with purpose.