An Electrochemical Biosensor Array for Rapid Detection of Alanine Aminotransferase and Aspartate Aminotransferase
-
- SONG Min-Jung
- School of Electrical Engineering, Korea University
-
- YUN Dong-Hwa
- Department of Chemical and Biological Engineering, Korea University
-
- HONG Suk-In
- Department of Chemical and Biological Engineering, Korea University
この論文をさがす
説明
An increment of alanine aminotransferase (ALT) or aspartate aminotransferase (AST) in human serum indicates an abnormal symptom of the liver. Hence, an electrochemical biosensor array that uses micro electro mechanical systems technology is required for rapid and integrated measurement of ALT/AST. Here we describe a biosensor array consisting of two glutamate sensors. It turned out that porous silicon layers formed on each working electrode were useful to increase the effective surface area. This biosensor array was constructed with platinum electrodes and a polydimethylsiloxane (PDMS) microchannel. Electrodes in sampling wells minimized a cross-interference effect and permitted multiple sampling by immobilization with glutamate oxidase using a silanization technique. The device sensitivities derived from semi-logarithmic plots were 0.145 μA/(U/l) for ALT and 0.463 μA/(U/l) for AST over a range of 1.3 U/l to 250 U/l. Hene, this ALT/AST biosensor array can be applied in diagnostic and home use.
収録刊行物
-
- Bioscience, Biotechnology, and Biochemistry
-
Bioscience, Biotechnology, and Biochemistry 73 (3), 474-478, 2009
公益社団法人 日本農芸化学会
- Tweet
キーワード
詳細情報 詳細情報について
-
- CRID
- 1390282681456897152
-
- NII論文ID
- 10027537979
-
- NII書誌ID
- AA10824164
-
- ISSN
- 13476947
- 09168451
-
- NDL書誌ID
- 10198868
-
- 本文言語コード
- en
-
- データソース種別
-
- JaLC
- NDL
- Crossref
- CiNii Articles
-
- 抄録ライセンスフラグ
- 使用不可