Neon isotope measurement in the presence of helium.

書誌事項

タイトル別名
  • Neon Isotope Measurement in the Presenc

この論文をさがす

抄録

Air standard neon (3.7×10-9 cm3 STP or 1.7×10-13 mol of Ne) was repeatedly analyzed with a sector type mass spectrometer VG5400 under various partial pressures of helium with the 4He/20Ne ratio from 0.24 to 27500 (or total pressure in the mass spectrometer, P(MS) from 1×10-7 Pa to 4.3×10-4 Pa by ion gauge reading). The measured 20Ne/22Ne ratios were kept almost constant when 4He/20Ne 400 or P(MS) 7×10-6 Pa (by ion gauge reading), but increased by up to a factor of ~3% at higher partial pressure of helium. The present result demonstrates the importance of He-Ne separation for a precise measurement of neon isotopic ratios in samples with high He/Ne ratios, such as MORB (Mid Oceanic Ridge Basalt) glasses.

収録刊行物

  • 質量分析

    質量分析 37 (6), 325-330, 1989

    一般社団法人 日本質量分析学会

被引用文献 (1)*注記

もっと見る

詳細情報 詳細情報について

問題の指摘