- Awaya Kiyoshi
- Tokyo Institute of Technology
-
- Yokoyama Isamu
- Tokyo Institute of Technology
-
- Shirahata Atsushi
- Watanabe Instruments Corp.
Bibliographic Information
- Other Title
-
- 精密音流計の研究
- セイミツオンリュウケイ ノ ケンキュウ
Search this article
Abstract
A sound-energy flux-meter applicable for accurate measurements is discussed. Its microphone head consists of two pressure microphones of nearly same characteristics; the voltage e_v proportional to the particle velocity v is obtained by time integration of the difference in voltage of the two microphone outputs. As the sound-energy flux density P is defined by the time average of p×v, the sound energy flux-meter must necessarily contain a multiplier such as the electronic wattmeter. The authors used a phase sensitive detector of new type as the multiplier, to construct a instrument with the accuracy of 2-3 %.
Journal
-
- THE JOURNAL OF THE ACOUSTICAL SOCIETY OF JAPAN
-
THE JOURNAL OF THE ACOUSTICAL SOCIETY OF JAPAN 15 (3), 159-172, 1959
Acoustical Society of Japan
- Tweet
Details 詳細情報について
-
- CRID
- 1390282681540325504
-
- NII Article ID
- 110003107441
-
- NII Book ID
- AN00186234
-
- ISSN
- 24322040
- 03694232
-
- NDL BIB ID
- 9144054
-
- Text Lang
- ja
-
- Data Source
-
- JaLC
- NDL
- CiNii Articles
-
- Abstract License Flag
- Disallowed